Flowmeter with AIMS CO2 Chamber
Example Calculations for 20% Chamber Vol/min
Large Rat Cage
19” x 10” x 8” = 1520 in3.
1520 in3 x 0.20 = 304 in3 (flow rate wanted 304 in3/min).
304 in3/min divided by 61.024 = 4.98 l/min AIMS CO2 flowmeter setting.
AIMS CO2 Chamber (cylinder)
π x Radius2 x Height
3.14 x 4”2 x 7” = 352 in3.
352 in3 x 0.20 = 70 in3 (flow rate wanted 70 in3/min).
70 in3/min divided by 61.024 = 1.15 l/min AIMS CO2 flowmeter setting.
Tubing w/Coupler Kit
1/4" to 1/4" ID - 2ea
1/4" to 5/16" ID - 2ea
5/16" to 5/16" ID - 2ea
5/16" to 3/8" ID - 2ea